Description: Learn how to identify common Auto-Light symptoms, verify the integration start conditions, and decide whether the fault is within the operator's authority or must be escalated. Auto-Light is a Class 4 integration laser, so the host machine's safeguards are part of every diagnosis.
First Response
- Stop the marking command. If an unsafe condition exists, operate the host machine's emergency stop and follow the site's isolation procedure.
- Keep personnel outside the laser hazard area. Do not open the enclosure until laser emission has stopped and the system is in the approved safe state.
- Before resetting anything, record the AnyMarker message, host-controller message, job name, machine state, and the last action performed.
- Check the host safety chain: emergency-stop channels, guard or access monitoring, shutter circuits, acknowledgements, emission warning, and extraction readiness.
- Check whether the expected X1 status sequence is present: emergency-stop status, SHUTTER, LASER READY, LASER BUSY, and LASER END.
Correct only conditions covered by operator training. If the condition repeats, cannot be identified, or requires electrical, optical, or internal access, isolate the system and contact authorised service personnel.
Capture the Fault Before Resetting
| Record | Example |
| System and job | Auto-Light serial/model, AnyMarker version, job name, selected device and parameter set |
| What happened | Would not become ready, stopped during marking, no mark, weak mark, axis fault, or communication loss |
| Signal state | X1 emergency-stop, SHUTTER, LASER READY, LASER BUSY and LASER END; host guard and extraction states |
| Timing and evidence | Time, last successful cycle, recent change, error text, and safe photographs or screenshots |
Safety, Readiness, and Unexpected Stops
| Symptom | Likely cause | Safe check or action |
| System will not power on | Mains supply absent; rear main switch off; damaged cable; host supply or protection device open | Keep the system off. Check the approved supply, plug and cable externally. Do not open the rack. Escalate damaged wiring or repeated tripping. |
| No LASER READY | Emergency-stop or shutter chain not acknowledged; guard/access condition open; controller or laser fault; integration sequence incomplete | Read the host and AnyMarker messages. Verify the guard, dual emergency-stop and dual shutter states, then perform only the approved acknowledgement/reset sequence. |
| SHUTTER status missing | Protective enclosure open; shutter circuit interrupted; acknowledgement missing; safety module fault | Close and secure the host enclosure. Confirm the host safety module reports both shutter channels healthy. Do not bridge the circuit. |
| Start command has no effect | LASER READY absent; external START not reaching X1.14; job not ready; another start condition missing | Confirm LASER READY first. Check the host command sequence and AnyMarker job state. Pin-level testing is for qualified integration personnel only. |
| Laser stops during a cycle | A start condition was lost; external STOP at X1.15; guard, emergency-stop, shutter, extraction or controller fault | Leave the system stopped. Record which status changed first. Correct the initiating condition; do not repeatedly reset a recurring fault. |
| BUSY or END sequence abnormal | Job/control communication issue; external control timing; controller fault | Record LASER READY/BUSY/END timing and the AnyMarker message. Check the approved host sequence; escalate if the sequence differs from the validated integration. |
No Mark, Weak Mark, or Long Cycle
| Symptom | Possible cause | Safe check or action |
| No visible mark | Safety permission absent; wrong job/device; unsuitable material; parameters produce no visible response | Confirm READY and SHUTTER status, the selected Auto-Light device and job, workpiece material, focus, and approved parameter assignment. Test only on an approved sample. |
| Mark is too weak | Incorrect Z value in the job or axis menu | Compare the commanded Z position with the validated setup. If an installed Z-axis has lost reference, stop and follow the approved axis-reference procedure before refocusing. |
| Mark is too weak | Focus is above or below the workpiece | Check the approved focal distance, fixture seating and workpiece height. Refocus before changing laser energy. |
| Mark is too weak | Objective contaminated or damaged | Stop and isolate the equipment. Inspect only as permitted by the maintenance procedure. Do not touch or clean optics unless trained and authorised. |
| Mark is too weak | Incorrect or unqualified parameters | Reload the validated recipe. If adjustment is authorised, change one controlled variable at a time and verify the result on an approved sample. |
| Marking takes too long | Low speed, insufficient effective energy, overly fine hatch spacing, too many passes, or unsuitable font | Compare the job with the released process. Use an approved plotter font where appropriate. Requalify any change that affects quality or cycle time. |
| Mark shifts or clips | Workpiece moved; fixture incorrect; pilot position, field, axis position or job origin changed | Stop emission. Secure the workpiece, verify the fixture and job origin with the pilot laser, and confirm axis position before another test. |
AnyMarker or Controller Communication
- Record the message before closing AnyMarker or restarting the controller.
- Confirm the approved USB or Ethernet connection and the expected Auto-Light device selection without disconnecting cables during an active cycle.
- If communication repeatedly drops, stop production and have the workstation, cable, controller and network path checked by authorised personnel.
For software messages, device connection, and job handling, use the AnyMarker Designer XWiki.
X1 Safety Interface - Emergency-Stop Circuit
The X1 emergency-stop connection uses two external safety channels. Circuit 1 is connected through X1.1/X1.2 and circuit 2 through X1.3/X1.4. The external safety module must validate both channels; a discrepancy belongs to the integrated safety chain, not to normal AnyMarker job handling.

Example: dual emergency-stop safety circuits at X1.1-X1.4.
Emergency-stop acknowledgement is provided at X1.5/X1.6. At every system restart and after every emergency-stop actuation acknowledgement is required. If acknowledgement is not accepted, do not repeat the reset indefinitely; record the host safety state and escalate a recurring condition.

Example: emergency-stop acknowledgement at X1.5/X1.6.
X1 Safety Interface - Shutter Circuit
The shutter connection also uses two external safety channels: circuit 1 at X1.7/X1.8 and circuit 2 at X1.9/X1.10. Opening the protective enclosure must interrupt the shutter circuit. A closed and validated shutter circuit permits active mode; it does not by itself command a marking cycle.

Example: dual shutter safety circuits at X1.7-X1.10.
Shutter acknowledgement is provided at X1.11/X1.12. When SHUTTER status is missing, first verify the host guard and safety-module indications. A pin-level test or change to the acknowledgement circuit is an integration-service task.

Example: shutter acknowledgement at X1.11/X1.12.
X1 Production and Status I/O
After the safety circuits are valid, use the documented X1 command and status sequence to localise a control fault. Record which state is missing or which output changes first.
| Pin group | Documented function | Diagnostic use |
| X1.13-X1.15 | +24 V input supply; External START; External STOP | If START has no effect, confirm LASER READY before checking the host command. An active STOP or missing start condition can stop a cycle. |
| X1.16-X1.17 | Emergency-stop status; SHUTTER status | These show whether the safety chain and shutter permission have reached the expected state. |
| X1.18-X1.20 | LASER READY; LASER BUSY; LASER END | Use the transition order to distinguish a missing ready condition, a cycle that never started, or a cycle that did not complete normally. |
| X1.21-X1.24 | Inputs 4 and 5; Outputs 4 and 5 | Their assigned use depends on the delivered integration. Check the approved electrical documentation before interpreting them. |
| X1.25 | GND for outputs | Active outputs supply 24 V and are evaluated against this output ground. The documented maximum output current is 250 mA. |
Expected Diagnostic Sequence
- Verify both emergency-stop channels and the required acknowledgement.
- Verify both shutter channels, the guard condition, and shutter acknowledgement.
- Confirm Emergency-stop status and SHUTTER status before expecting LASER READY.
- With LASER READY present, check whether External START is received and whether LASER BUSY then LASER END follows the validated host sequence.
- If the sequence differs, record the states and stop. Do not change timing, wiring, or safety logic unless authorised for the integration.
X2 Axis Interface
X2 is the 25-pin Sub-D axis interface. Direction and step outputs for X, Y, Z, and R are grouped with zero/reference signals. Use this pinout to identify the affected axis and whether the symptom concerns commanded motion or the reference input; do not use it as an energized probing guide.
| Pin | Direction | Description |
|---|---|---|
| 1 | Output | Direction X |
| 2 | Output | Step X |
| 3 | Output | Direction Y |
| 4 | Output | Step Y |
| 5 | Output | Direction Z |
| 6 | Output | Step Z |
| 7 | Output | Direction R |
| 8 | Output | Step R |
| 9 | Output | Zero X |
| 10 | - | Not connected |
| 11 | Input | Zero Y |
| 12 | - | Not connected |
| 13 | Input | Zero Z |
| 14 | - | Not connected |
| 15 | Input | Zero R |
| 16 | - | Not connected |
| 17 | Output | GND |
| 18 | - | Not connected |
| 19 | Output | +5V supply |
| 20 | - | Not connected |
| 21 | - | Not connected |
| 22 | - | Not connected |
| 23 | - | Not connected |
| 24 | - | Not connected |
| 25 | - | Not connected |
Diagnosing an X2 Axis Fault
| Axis | Direction output | Step output | Zero/reference signal |
| X | X2.1 | X2.2 | X2.9 |
| Y | X2.3 | X2.4 | X2.11 |
| Z | X2.5 | X2.6 | X2.13 |
| R | X2.7 | X2.8 | X2.15 |
Use the Signal Groups to Localise the Fault
- No commanded movement: identify the relevant direction/step pair - X2 pins 1-2 for X, 3-4 for Y, 5-6 for Z, or 7-8 for R - then check the approved axis configuration and host mechanics.
- Axis will not reference: inspect the corresponding zero signal - X2.9 for X, X2.11 for Y, X2.13 for Z, or X2.15 for R - using the approved service procedure.
- Unexpected direction, travel, repeated reference failure, or mechanical obstruction: stop movement and isolate the system. Do not continue jogging the axis to overcome the condition.
Operator Checkpoints on the Laser Rack

- Check the rear main switch, front laser key switch, status indicator, and approved host/AnyMarker connection.
- Keep the cooling vents unobstructed. Stop operation if airflow is restricted, heat accumulates, or an abnormal smell or sound is present.
- Visually inspect accessible cables and connectors without opening covers or disturbing the validated wiring. Isolate and escalate damage, looseness, overheating, or contamination.
Stop and Escalate
Stop troubleshooting, isolate the equipment, and contact authorised service personnel when:
- a safety channel, shutter circuit, guard function, acknowledgement or emission warning cannot be verified;
- the same fault returns after one approved reset, or LASER READY/BUSY/END behaves unpredictably;
- internal electrical, optical, cooling, laser-source, controller, X1/X2 wiring, or mechanical-axis access is required;
- an optic, cable, connector, enclosure, laser-protection window, fixture, or extraction component is damaged; or
- the cause is uncertain or the corrective action is outside the person's training and authority.
Information for Schilling Service
Provide the serial number and exact Auto-Light model, AnyMarker version and job name, all host and laser messages, recorded X1 states, the event sequence, recent changes, and safe photographs or screenshots.