Last modified by jadomatis on 2026/09/08 07:54

From version 2.3
edited by jadomatis
on 2026/09/08 07:41
Change comment: There is no comment for this version
To version 3.1
edited by jadomatis
on 2026/09/08 07:54
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -1,5 +1,7 @@
1 1  **Description**: Learn to identify the standard Auto-Light mechanical, optical, laser-source, electrical, and environmental data. Auto-Light is a modular Class 4 integration laser available in F20, F30, F50, M20, and M30 variants.
2 2  
3 +{{toc/}}
4 +
3 3  (% class="box warningmessage" %)
4 4  (((
5 5  **Configuration control: **The values below describe the standard configurations. The nameplate, order confirmation, delivered drawings, and approved electrical documentation for the specific system take precedence. Options, replacement components, or a different lens can change relevant values.
... ... @@ -10,7 +10,7 @@
10 10  = System Identification =
11 11  
12 12  (% class="table-condensed table-striped" %)
13 -|**Property**|**Manual specification**
15 +|**Property**|**Specification**
14 14  |System type|Integration laser
15 15  |Laser classification|Laser Class 4
16 16  |Protection rating|IP21
... ... @@ -27,8 +27,6 @@
27 27  
28 28  = Laser-Source Variants =
29 29  
30 -The F models use a pulsed fiber laser. The M models use a pulsed MOPA fiber laser. Values in both tables are reproduced from the manual's data sheet.
31 -
32 32  (% class="table-condensed table-striped" %)
33 33  |**Model**|**Laser source**|**Nominal power**|**Pulse energy**|**Beam quality**
34 34  |F20|Pulsed fiber|20 W|1 mJ|M2 < 1.5
... ... @@ -68,4 +68,49 @@
68 68  |Laser-rack dimensions (L x W x H)|mm|550 x 441 x 162
69 69  |Laser-head dimensions|mm|367 x 131 x 131
70 70  
71 +(% class="box warningmessage" %)
72 +(((
73 + **Standard optics: **The 176 mm working distance and 110 x 110 mm field apply to the listed 160 mm F-Theta lens. Confirm the actual lens before setting the fixture height, enclosure geometry, extraction nozzle, or marking boundary.
74 +)))
75 +
76 +
77 +
78 += Operating and Installation Conditions =
79 +
80 +(% class="table-condensed table-striped" %)
81 +|**Condition**|**Requirement**
82 +|Ambient temperature|+15 degrees C to +35 degrees C
83 +|Relative humidity|No higher than 90%, non-condensing
84 +|After a major temperature change|Allow the system to return to room temperature before commissioning
85 +|Cooling-air openings|Keep approximately 15 cm clear and do not obstruct ventilation slots or filter mats
86 +|Operating environment|Avoid mechanical loads, vibration, impact, dust accumulation, and heat build-up
87 +|Process emissions|Provide extraction suitable for the material, fumes, vapors, dust, and particles produced
88 +
89 +
90 +
91 += Interfaces and Supply - Overview =
92 +
93 +(% class="table-condensed table-striped" %)
94 +|**Connection**|**Description**|**Purpose**
95 +|X0|230 V / 600 W, including main switch|Mains-supply interface
96 +|X1|25-pin Sub-D socket|Dual emergency-stop and shutter circuits, acknowledgements, external START/STOP, and 24 V status/production I/O
97 +|X1 outputs|24 V when active; Imax 250 mA|External status and assigned production outputs referenced to X1.25 GND
98 +|X2|25-pin Sub-D plug|Direction, step, and zero signals for X, Y, Z, and R axes, with documented +5 V and GND
99 +|Rack data connections|USB, Ethernet, and HDMI|Workstation, network, and display connections according to the delivered integration
100 +
101 +(% class="box errormessage" %)
102 +(((
103 + **Safety interface: **X1 is part of the integrated safety and control boundary. A connector rating or pin description is not a complete safety design. Do not bridge, modify, or test safety circuits unless the system is isolated and the work is performed under the approved procedure by authorised integration personnel.
104 +)))
105 +
106 +
107 +
108 += Verification Before Use =
109 +
110 +* Match the model designation and serial number to the nameplate and order documentation.
111 +* Confirm the installed lens, focal length, working distance, marking field, and wavelength before fixing the mechanical design or laser-protection concept.
112 +* Verify X0 supply data and all X1/X2 assignments against the delivered electrical drawings; do not rely on this summary for wiring.
113 +* Confirm that the host machine provides and validates the required enclosure, access monitoring, emergency stop, shutter control, emission warning, extraction, and other site-specific safeguards.
114 +* Record approved deviations or options in the machine technical file and update the applicable KB article when the as-built configuration changes.
115 +
71 71