1 **Description**: Learn how to identify the Auto-Light X0 mains connection, X1 safety and production interface, and X2 axis interface. The pin assignments and diagrams are reproduced from the supplied operating manual in English.
2
3 (% class="box errormessage" %)
4 (((
5 **Qualified integration work only: **The manual diagrams are connection examples, not a complete safety design or permission to wire an energised system. An appropriately qualified integrator must design, calculate, isolate, wire, validate, and document the complete safety-related control system for the delivered machine.
6 )))
7
8 ----
9
10 = Connection Locations =
11
12 X0 mains power, X1 safety and production I/O, and X2 axis control are located at the rear of the laser rack. USB, Ethernet, and HDMI connections support the delivered workstation, network, and display arrangement.
13
14 [[image:1788847354911-289.png]]
15
16
17 = X0 Mains Supply =
18
19 The manual identifies X0 as a 230 V / 600 W mains-supply interface including the main switch. Before connection, verify the actual nameplate, protective conductor, supply frequency, protective measures, disconnecting means, cable, plug, and site electrical documentation.
20
21 (% class="box warningmessage" %)
22 (((
23 **Keep X0 separate from model consumption data: **The X0 declaration is the documented supply-interface value. The lower power-consumption figures listed for individual laser-source variants in Technical Specifications do not replace the delivered supply and protection requirements.
24 )))
25
26
27 = X1 Laser Interface - Pin Assignment =
28
29 (% class="table-condensed table-striped" %)
30 |**Pin(s)**|**Signal**|**Manual description / use**
31 |1-2|Emergency-stop circuit 1|External wiring through a safety module required
32 |3-4|Emergency-stop circuit 2|External wiring through a safety module required
33 |5-6|Emergency-stop acknowledgement|Acknowledgement contact
34 |7-8|Shutter circuit 1|External wiring through a safety module required
35 |9-10|Shutter circuit 2|External wiring through a safety module required
36 |11-12|Shutter acknowledgement|Acknowledgement contact
37 |13|+24 V|Supply voltage for inputs
38 |14|External START|24 V input
39 |15|External STOP|24 V input
40 |16|Emergency-stop status|24 V output
41 |17|SHUTTER status|24 V output
42 |18|LASER READY status|24 V output
43 |19|LASER BUSY status|24 V output
44 |20|LASER END status|24 V output
45 |21-22|Inputs 4 and 5|24 V inputs
46 |23-24|Outputs 4 and 5|24 V outputs
47 |25|GND|Ground for outputs
48
49
50
51 = X1 Emergency-Stop Connection Examples =
52
53 The manual example routes two external emergency-stop channels through a safety module. Circuit 1 uses X1.1/X1.2 and circuit 2 uses X1.3/X1.4. Both channels belong to the integrated safety system and must be assessed and validated as part of the host machine.
54
55 [[**Example: dual emergency-stop circuits at X1.1-X1.4.**>>image:1788848413333-368.png]]
56
57
58 Emergency-stop acknowledgement is provided at X1.5/X1.6. TAcknowledgement is required at every restart of the system and after every operation of the emergency stop.
59
60 [[**Example: emergency-stop acknowledgement at X1.5/X1.6.**>>image:1788848413336-918.png]]