3 3  Confirm the following before applying power:
4 4  
5 -* Electrical
5 +* Electrical installation is correct; mains voltage and frequency match the rating place. Main switch is OFF.
6 +* Protective housing and door interlocks are intact; laser viewing window shows now damage.
7 +* Optics are clean; remove any transport/lens cover only now. Avoid touching optics with bare hands.
8 +* Ambient conditions: +15C to +35C. RH <90% non condensing; system acclimatized if temperature changed during transport.
9 +* Airflow at 19" drawers is unobstructed; filter mats are clean to prevent heat build-up.
10 +* Operator is familiar with laser safety requirements; responsibilities are defined.
11 +Note: The Multi-Light contains a Class 4 source operated as Class 1 in the closed cabin. Never bypass safety devices.
12 +
13 +
14 +
15 += Electrical Start-Up =
16 +
17 +1. Connect the system to the specified mains supply. Ensure extraction is running.
18 +1. Set the mains switch to ON. Internal electronics initialize; status indicators light as described on the operator panel.
19 +1. Power on the control PC (or workstation) and peripherals needed for operation.
20 +
21 +
22 +
23 += Software and Device Connection =
24 +
25 +1. Launch the AnyMarker Designer software.
26 +1. Verify communication with the laser controller in the software (device connected, parameters readable).
27 +If connection fails, switch OFF, check cabling and restart.
28 +
29 +
30 +
31 += Initial Checks With Pilot Laser =
32 +
33 +* Load or create a test layout in AnyMarker Designer.
34 +* Enable the pilot laser to project contours on the workpiece; verify orientation and placement without laser emission. Keep the door closed.
35 +
36 +
37 +
38 += Z-Axis Focusing =
39 +
40 +1. Place the workpiece/fixture in the work chamber and close the manual lift door.
41 +1. Use the automatic Z-axis to set focus to the surface (per lens focal distance). Follow the on-screen/readout guidance for focal height.
42 +1. Confirm best focus by pilot outline sharpness or focus gauge method recommended in the manual.
43 +
44 +
45 +
46 += Parameter Verification =
47 +
48 +* Select a verified marking parameter set (power, speed, frequency, fill). Start with conservative energy to avoid burning or glare.
49 +* Ensure extraction is active to remove gases, vapors, aerosols, and dust during test marking.
50 +
51 +
52 +
53 += First Test Mark =
54 +
55 +1. With the cabin closed and interlocks engaged, execute a short test mark on a sample of the target material.
56 +1. Observe quality and adjust parameters (e.g., reduce power/increase speed) as needed. Multiple repetitions may be used for depth, but monitor temperature to reduce fire risk.
57 +1. If unexpected behavior occurs, press STOP/EMERGENCY-STOP, investigate messages in Notifications, and correct the cause before resuming.
58 +
59 +
60 +
61 +