5 5  * Electrical installation is correct; mains voltage and frequency match the rating place. Main switch is OFF.
6 6  * Protective housing and door interlocks are intact; laser viewing window shows now damage.
7 7  * Optics are clean; remove any transport/lens cover only now. Avoid touching optics with bare hands.
8 -* Ambient conditions: +15C to +35C. RH <90% non condensing; system acclimatized if temperature changed during transport.
9 -* Airflow at 19" drawers is unobstructed; filter mats are clean to prevent heat build-up.
10 -* Operator is familiar with laser safety requirements; responsibilities are defined.
11 -Note: The Multi-Light contains a Class 4 source operated as Class 1 in the closed cabin. Never bypass safety devices.
12 -
13 -
14 -
15 -= Electrical Start-Up =
16 -
17 -1. Connect the system to the specified mains supply. Ensure extraction is running.
18 -1. Set the mains switch to ON. Internal electronics initialize; status indicators light as described on the operator panel.
19 -1. Power on the control PC (or workstation) and peripherals needed for operation.
20 -
21 -
22 -
23 -= Software and Device Connection =
24 -
25 -1. Launch the AnyMarker Designer software.
26 -1. Verify communication with the laser controller in the software (device connected, parameters readable).
27 -If connection fails, switch OFF, check cabling and restart.
28 -
29 -
30 -
31 -= Initial Checks With Pilot Laser =
32 -
33 -* Load or create a test layout in AnyMarker Designer.
34 -* Enable the pilot laser to project contours on the workpiece; verify orientation and placement without laser emission. Keep the door closed.
35 -
36 -
37 -