31 += Initial Checks With Pilot Laser =
32 +
33 +* Load or create a test layout in AnyMarker Designer.
34 +* Enable the pilot laser to project contours on the workpiece; verify orientation and placement without laser emission. Keep the door closed.
35 +
36 +
37 +
38 += Z-Axis Focusing =
39 +
40 +1. Place the workpiece/fixture in the work chamber and close the manual lift door.
41 +1. Use the automatic Z-axis to set focus to the surface (per lens focal distance). Follow the on-screen/readout guidance for focal height.
42 +1. Confirm best focus by pilot outline sharpness or focus gauge method recommended in the manual.
43 +
44 +
45 +
46 += Parameter Verification =
47 +
48 +* Select a verified marking parameter set (power, speed, frequency, fill). Start with conservative energy to avoid burning or glare.
49 +* Ensure extraction is active to remove gases, vapors, aerosols, and dust during test marking.
50 +
51 +
52 +
53 += First Test Mark =
54 +
55 +1. With the cabin closed and interlocks engaged, execute a short test mark on a sample of the target material.
56 +1. Observe quality and adjust parameters (e.g., reduce power/increase speed) as needed. Multiple repetitions may be used for depth, but monitor temperature to reduce fire risk.
57 +1. If unexpected behavior occurs, press STOP/EMERGENCY-STOP, investigate messages in Notifications, and correct the cause before resuming.
58 +