1 1  **Description**: Learn how to identify common Auto-Light symptoms, verify the integration start conditions, and decide whether the fault is within the operator's authority or must be escalated. Auto-Light is a Class 4 integration laser, so the host machine's safeguards are part of every diagnosis.
2 2  
3 +{{toc/}}
4 +
3 3  (% class="box errormessage" %)
4 4  (((
5 5  **Safety first: **Never bypass a guard, emergency-stop channel, shutter circuit, acknowledgement circuit, emission warning, extraction interlock, or protective cover to test the laser. Maintenance and repair may be performed only by Schilling Marking Systems or authorised personnel in accordance with the site's safety procedure.
... ... @@ -64,6 +64,7 @@
64 64  
65 65  
66 66  
69 +
67 67  = AnyMarker or Controller Communication =
68 68  
69 69  * Record the message before closing AnyMarker or restarting the controller.
... ... @@ -70,7 +70,7 @@
70 70  * Confirm the approved USB or Ethernet connection and the expected Auto-Light device selection without disconnecting cables during an active cycle.
71 71  * If communication repeatedly drops, stop production and have the workstation, cable, controller and network path checked by authorised personnel.
72 72  
73 -For software messages, device connection, and job handling, use the **[[AnyMarker Designer XWiki>>url:https://docs2.anymarker.de]]**.
76 +For software messages, device connection, and job handling, use the **[[AnyMarker Designer XWiki>>url:https://docs2.anymarker.de]]**. 
74 74  
75 75  
76 76  = X1 Safety Interface - Emergency-Stop Circuit =
... ... @@ -81,13 +81,14 @@
81 81  )))
82 82  
83 83  (% style="font-size:14px" %) (%%)The X1 emergency-stop connection uses two external safety channels. Circuit 1 is connected through X1.1/X1.2 and circuit 2 through X1.3/X1.4. The external safety module must validate both channels; a discrepancy belongs to the integrated safety chain, not to normal AnyMarker job handling.
84 -\\
87 +
85 85  
86 86  [[**Example: dual emergency-stop safety circuits at X1.1-X1.4.**>>image:1788775595167-399.png]]
87 87  
88 88  
92 +
89 89  Emergency-stop acknowledgement is provided at X1.5/X1.6. At every system restart  and after every emergency-stop actuation acknowledgement is required. If acknowledgement is not accepted, do not repeat the reset indefinitely; record the host safety state and escalate a recurring condition.
90 -\\
94 +
91 91  
92 92  [[**Example: emergency-stop acknowledgement at X1.5/X1.6.**>>image:1788775764610-963.png]]
93 93  
... ... @@ -101,7 +101,7 @@
101 101  
102 102  
103 103  Shutter acknowledgement is provided at X1.11/X1.12. When SHUTTER status is missing, first verify the host guard and safety-module indications. A pin-level test or change to the acknowledgement circuit is an integration-service task.
104 -\\
108 +
105 105  
106 106  [[**Example: shutter acknowledgement at X1.11/X1.12.**>>image:1788776541211-163.png]]
107 107  
... ... @@ -139,6 +139,7 @@
139 139  
140 140  
141 141  
146 +
142 142  = X2 Axis Interface =
143 143  
144 144  X2 is the 25-pin Sub-D axis interface. Direction and step outputs for X, Y, Z, and R are grouped with zero/reference signals. Use this pinout to identify the affected axis and whether the symptom concerns commanded motion or the reference input; do not use it as an energized probing guide.
... ... @@ -202,6 +202,7 @@
202 202  
203 203  
204 204  
210 +
205 205  = Operator Checkpoints on the Laser Rack =
206 206  
207 207  
... ... @@ -237,3 +237,5 @@
237 237  **service@schilling-marking.de**  |  **www.schilling-marking.de**
238 238  In Grubenaecker 1, 78532 Tuttlingen, Germany  |  **Fax +49 (07461) 9472-20**
239 239  )))
246 +
247 +