... ... @@ -97,50 +97,7 @@
97 97  The shutter connection also uses two external safety channels: circuit 1 at X1.7/X1.8 and circuit 2 at X1.9/X1.10. Opening the protective enclosure must interrupt the shutter circuit. A closed and validated shutter circuit permits active mode; it does not by itself command a marking cycle.
98 98  
99 99  
100 -[[**Example: dual shutter safety circuits at X1.7-X1.10.**>>image:1788776328447-524.png]]
101 -
102 -
103 -Shutter acknowledgement is provided at X1.11/X1.12. When SHUTTER status is missing, first verify the host guard and safety-module indications. A pin-level test or change to the acknowledgement circuit is an integration-service task.
100 +[[image:1788776328447-524.png]]
104 104  \\
105 105  
106 -[[**Example: shutter acknowledgement at X1.11/X1.12.**>>image:1788776541211-163.png]]
107 -
108 -(% class="box warningmessage" %)
109 -(((
110 -**Diagnostic meaning: **If either emergency-stop or shutter channel is open, inconsistent, or not acknowledged, LASER READY may remain absent. Never bridge a channel to prove this diagnosis; use the host safety-module indications and approved electrical test procedure.
111 -)))
112 -
113 -
114 -= X1 Production and Status I/O =
115 -
116 -After the safety circuits are valid, use the documented X1 command and status sequence to localise a control fault. Record which state is missing or which output changes first.
117 -
118 -|**Pin group**|**Documented function**|**Diagnostic use**
119 -|X1.13-X1.15|+24 V input supply; External START; External STOP|If START has no effect, confirm LASER READY before checking the host command. An active STOP or missing start condition can stop a cycle.
120 -|X1.16-X1.17|Emergency-stop status; SHUTTER status|These show whether the safety chain and shutter permission have reached the expected state.
121 -|X1.18-X1.20|LASER READY; LASER BUSY; LASER END|Use the transition order to distinguish a missing ready condition, a cycle that never started, or a cycle that did not complete normally.
122 -|X1.21-X1.24|Inputs 4 and 5; Outputs 4 and 5|Their assigned use depends on the delivered integration. Check the approved electrical documentation before interpreting them.
123 -|X1.25|GND for outputs|Active outputs supply 24 V and are evaluated against this output ground. The documented maximum output current is 250 mA.
124 -
125 -
126 -
127 -== Expected Diagnostic Sequence ==
128 -
129 -1. Verify both emergency-stop channels and the required acknowledgement.
130 -1. Verify both shutter channels, the guard condition, and shutter acknowledgement.
131 -1. Confirm Emergency-stop status and SHUTTER status before expecting LASER READY.
132 -1. With LASER READY present, check whether External START is received and whether LASER BUSY then LASER END follows the validated host sequence.
133 -1. If the sequence differs, record the states and stop. Do not change timing, wiring, or safety logic unless authorised for the integration.
134 -
135 -(% class="box warningmessage" %)
136 -(((
137 -**Status outputs are evidence, not permission to bypass: **A missing signal helps locate the fault but does not authorise forcing an input or output. Troubleshoot the host safety system, Auto-Light interface, and controller as the validated integrated machine.
138 -)))
139 -
140 -
141 -
142 -= X2 Axis Interface =
143 -
144 -X2 is the 25-pin Sub-D axis interface. Direction and step outputs for X, Y, Z, and R are grouped with zero/reference signals. Use this pinout to identify the affected axis and whether the symptom concerns commanded motion or the reference input; do not use it as an energized probing guide.
145 -
146 146