49 -= No Mark, Weak Mark, or Long Cycle =
50 -
51 -|**Symptom**|**Possible cause**|**Safe check or action**
52 -|No visible mark|Safety permission absent; wrong job/device; unsuitable material; parameters produce no visible response|Confirm READY and SHUTTER status, the selected Auto-Light device and job, workpiece material, focus, and approved parameter assignment. Test only on an approved sample.
53 -|Mark is too weak|Incorrect Z value in the job or axis menu|Compare the commanded Z position with the validated setup. If an installed Z-axis has lost reference, stop and follow the approved axis-reference procedure before refocusing.
54 -|Mark is too weak|Focus is above or below the workpiece|Check the approved focal distance, fixture seating and workpiece height. Refocus before changing laser energy.
55 -|Mark is too weak|Objective contaminated or damaged|Stop and isolate the equipment. Inspect only as permitted by the maintenance procedure. Do not touch or clean optics unless trained and authorised.
56 -|Mark is too weak|Incorrect or unqualified parameters|Reload the validated recipe. If adjustment is authorised, change one controlled variable at a time and verify the result on an approved sample.
57 -|Marking takes too long|Low speed, insufficient effective energy, overly fine hatch spacing, too many passes, or unsuitable font|Compare the job with the released process. Use an approved plotter font where appropriate. Requalify any change that affects quality or cycle time.
58 -|Mark shifts or clips|Workpiece moved; fixture incorrect; pilot position, field, axis position or job origin changed|Stop emission. Secure the workpiece, verify the fixture and job origin with the pilot laser, and confirm axis position before another test.