10 -= Auto-Light System Overview =
11 -
12 -* The laser rack contains the fibre-laser source and control hardware.
13 -* The marking head directs and scans the beam through the installed F-Theta lens.
14 -* The manual documents a 3 m fibre arrangement between the source and marking head.
15 -* A pilot laser supports workpiece positioning before production-laser emission.
16 -* An electrically controlled Z-axis establishes the required working distance in the documented configuration.
17 -* X1 connects external safety and production signals; X2 supports integrated X/Y/Z/R axis signals.
18 -
19 -[[**Auto-Light laser rack and marking head.**>>image:1788852299021-764.png]]
20 -
21 -
22 -= Operator and Connection Panels =
23 -
24 -[[**Auto-Light front and rear panels. **>>image:1788852351064-324.png]]
25 -
26 -
27 -(% class="table-condensed table-striped" style="width:840px" %)
28 -|**Location**|(% style="width:628px" %)**Documented items**
29 -|Front panel|(% style="width:628px" %)Shutter / laser key switch; status indicator; USB interface; cooling-air openings.
30 -|Rear panel|(% style="width:628px" %)HDMI; Ethernet; USB; X1 interface; X2 interface; main switch; mains input.
31 -
32 -(% class="box warningmessage" %)
33 -(((
34 - **Ventilation clearance: **Keep approximately 15 cm clear at the cooling-air openings. Blocked airflow can cause heat accumulation and equipment damage.
35 -)))
36 -
37 -
38 -
39 -= Protective Devices =
40 -
41 -Only the main switch and shutter as protective devices on the Auto-Light. Other controls and interfaces support the safety concept but do not, by themselves, complete it.
42 -
43 -|**Device**|**Safety function and limitation**|**Required condition**
44 -|Main switch|Controls the rack's electrical supply for normal start and shutdown. It is not a substitute for the site's required isolation procedure during electrical or service work.|Must be accessible, undamaged, and used according to the approved start, shutdown, and isolation procedure.
45 -|Shutter|Interrupts the laser beam when the integrated safety logic removes shutter permission. A closed shutter circuit enables active mode; it does not issue a marking command.|Both external shutter channels, acknowledgement, guard logic, and beam-interruption response must be validated as part of the host machine.
46 -
47 47  