Last modified by jadomatis on 2026/09/07 09:02

From version 4.5
edited by jadomatis
on 2026/09/07 08:37
Change comment: There is no comment for this version
To version 4.7
edited by jadomatis
on 2026/09/07 08:39
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -77,4 +77,27 @@
77 77  
78 78  
79 79  
80 -
80 += Shutdown =
81 +
82 +1. Finish or stop the current job and confirm that laser emission has ended.
83 +1. Save any required production data and close AnyMarker Designer in accordance with the site procedure.
84 +1. Set the rear main switch to 0 (OFF).
85 +1. The operating manual permits the front key switch to remain at I or be returned to 0. Follow the site's approved key-control and access procedure.
86 +1. Keep extraction running until remaining fumes and dust have been removed, then shut it down according to the extraction-system instructions.
87 +1. Leave the processing area clean and secure. Switch off the main switch whenever the machine is not in use.
88 +
89 +
90 +
91 += If Something Does Not Work =
92 +
93 +(% class="table-condensed table-striped" style="width:1070px" %)
94 +|(% style="width:309px" %)**Symptom**|(% style="width:760px" %)**Check**
95 +|(% style="width:309px" %)No ready signal|(% style="width:760px" %)Host safety acknowledgement, emergency-stop circuits, shutter circuits, enclosure status, and active faults.
96 +|(% style="width:309px" %)No marking|(% style="width:760px" %)Safety permission, shutter status, focus, job parameters, connection to the controller, and material suitability.
97 +|(% style="width:309px" %)Weak marking|(% style="width:760px" %)Z-axis position, focal distance, objective cleanliness, protective glass, and the approved parameter set.
98 +|(% style="width:309px" %)Unexpected stop|(% style="width:760px" %)Treat the loss of a start condition as a fault. Secure and isolate the system before investigation.
99 +
100 +(% class="box warningmessage" %)
101 +(((
102 +**Unresolved fault: **Switch off the system and contact Schilling Service or authorised maintenance personnel. Repairs must not be attempted by unqualified personnel
103 +)))