Wiki source code of Spare Parts and Consumables
Hide last authors
| author | version | line-number | content |
|---|---|---|---|
| |
1.1 | 1 | Learn about the typical wear parts and consumables for Multi-Light workstation and how to replace them. Always use original parts and follow the maintenance instructions in the manual. |
| 2 | |||
| 3 | |||
| 4 | = Filter Mat = | ||
| 5 | |||
| |
1.3 | 6 | What it does: Protects the electronics from dust; keeps cooling airflow stable. |
| |
1.4 | 7 | Location: Behind the front ventilation slots of the laser plug-in unit or the PC. |
| |
1.5 | 8 | When to replace: Inspect regularly; replace when soiled to maintain proper cooling. |
| 9 | |||
| 10 | |||
| 11 | = Extraction System Filters = | ||
| 12 | |||
| |
1.6 | 13 | What do they do: Remove fumes, vapors, aerosols, and dusts generated during marking; protect personnel and optics. |
| |
1.7 | 14 | When to replace: At least annually check the main filter; replace if contaminated. Frequency depends on workload/material. |
| |
1.9 | 15 | Tip: Using an extraction unit is highly recommended for plastics and larger material removal (e.g., anodized/painted/coated parts). Systems are available in multiple sizes and can be retrofitted. |
| 16 | |||
| 17 | |||
| |
1.10 | 18 | = Laser Protection Window = |
| 19 | |||
| 20 | What it does: Provides certified protection at the laser wavelength and allows visual monitoring. | ||
| |
1.11 | 21 | When to replace: If scratched, cracked, or clouded - stop operation and replace before further use. Clean only with approved methods. |
| 22 | |||
| 23 | |||
| |
1.12 | 24 | = Perimeter Seals/Cladding Fasteners = |
| 25 | |||
| |
1.13 | 26 | What they do: Ensure the protective cabin remains Class 1 tight (no radiation leakage). |
| 27 | When to replace: If missing, damaged, or loose. Operation with compromised seals/cladding is not permitted. | ||
| 28 | |||
| 29 | |||
| 30 | = Optics Care Material = | ||
| 31 | |||
| 32 | What to stock: | ||
| 33 | |||
| 34 | * Approved lens cleaning wipes/lint-free tissues | ||
| 35 | * High-purity alcohol (>98%) for optics cleaning | ||
| 36 | * Small hand air bulb (no compressed air) | ||
| 37 | When to use: As needed when contamination is evident; follow gentle, center-to-edge cleaning technique. | ||
| 38 | |||
| 39 | |||
| 40 | |||
| 41 | = Lenses/Objectives = | ||
| 42 | |||
| 43 | What they do: Change field/spot size (e.g., F-Theta f=160 mm for 110 x 110 mm field) | ||
| 44 | When to replace/add: When application requires a different field or when an optic is irreparably damaged (scratched). Re-focus after any lens change. | ||
| 45 | |||
| 46 | |||
| |
1.14 | 47 | = Service Items (performed by Schilling/authorized personnel) = |
| 48 | |||